Ventilation/Perfusion Lung Imaging
Purpose:
- Primary reason - to determine if the patient has pulmonary embolism/i (PE)
- Can evaluate Chronic Obstruction Pulmonary Disease (COPD)
- Can evaluate lung function for the purpose of pneumonectomy (covered next semester)
To evaluate PE, it requires two lung procedures: Ventilation and Perfusion
- Ventilation - 133Xe (gas) - Covered next semester
- Dose - 10 to 20 mCi
- 5.3 day T1/2
- 80 keV gamma
- Nobel gas
- Allows for separation of peaks when imaging with 99mTcMAA
- Ventilation - 99mTcDTPA (aerosol)
- Dose
- 35 mCi if dose before perfusion (preferred method)
- 50 - 60 mCi if dosed after perfusion
- 6.02 hr T1/2
- 140 keV gamma
- Nebulized droplets that are less than 2.0 μm reach the alveoli
- It does not allow for separation of peak when imaging the perfusion study
- Ventilation can be done with radioactive gas - 81mKr or 133Xe and will be discussed next semester
- Perfusion - 99mTcMAA - see previous lecture
- Dose - 4 to 6 mCi
- Small particles used and is trapped by capillary blockage
- Size 10 - 90 microns at 90% with nothing greater than 150 microns at 10%
- Trapped in the arterioles of the lung
- Fewer than one arteriole is blocked for every 1000
Disease
- Pulmonary Embolism (PE)
-
Originates from deep vein thrombi (DVT's) found in legs
-
DVT's break off, move through the venous system and become trapped in the lungs
-
When injecting MAA particles, the particles cannot pass the area that has a PE
-
Results in a cold defect when imaged
with MAA
- Ventilation shows normal uptake since the activity is in the bronchioles (PE does not trap the airway)
- Key point - PE appears cold in perfusion and hot in ventilation (mismatched defect)
-
Symptoms of PE
-
Shortness of breath/air (SOA)
-
Sudden onset of chest pain
-
Dyspnea
-
Cardiac arrhythmias
-
Drop in the patient's PO2 level (normal 90-110)
- Aerosol Ventilation
- Aerosol system
- Approximately 35 mCi of 99mTcDTPA is injected into the nebulizer (refer to diagram)
- O2 enters the nebulizer converting liquid (99m TcDTPA) into small liquid droplets
- The patient inhales and droplets coating the inside of the lungs
- The patient exhales and the remaining droplets and water vapor are trapped by the filter
- Note the one-way value in the diagram
- The system is shielded in a blue bucket
- Example of another aerosol system
- Below is a display of how a patient is dosed if he/she is on a ventilator. Extra tubing is added at the end of the system referred to as the "tail"
- Dosing the aerosol
- Usually, the patient breaths on this system five minutes
- After the ventilation process images are then taken at 150k counts per view or five minutes
- View may include: ANT, POST, LPO, RPO, LAO, RAO, LLAT, and RLAT
- This procedure is usually done before the Perfusion images (99mTcMAA)
- All images that are taken must match all the perfusion images taken (at the exact same angle)
- PE will show a mismatch defect(s) - hot in ventilation and cold in perfusion
- Usually, PE defects are segmental or subsegmental
- Usually, PE is multiple in occurrence
- COPD will show matched defect(s) - cold in ventilation and cold in perfusion
- Severe COPD will show clumping of the droplets in the lungs - "hot grapes" and may cause the study to be mis-read
Aerosol Ventilation Procedure
Xenon Ventilation Procedure
Return to the beginning of the document
Return to the Table of Contents
6/21