AtomLab 100 plus - Dose Calibrator
QC and Isotope setting
The dose calibrator (DC) power should always be on, however, should you need to turn it on the switch can be found in the back
QC should be done first thing in the morning, every morning, when you are open to receive patients
Self-Test is initiated when the power is turned on
However, if the power is already on select the Test key (blue box)
This test evaluates the over electrons and operation of the DC
Electrometer sensitivity chamber bias supply, software checksum, and display settings
The following sequence is noted, once the Test key is selected
The main LED will show " 8.8.8.8" followed by the smaller LED (yellow box) displaying the same series of 8s
Next all green buttons will be eliminated
Ending with the words, "high" and "elec" displayed in the yellow box
Once the test is over the main LED will display the background counts within the chamber
The test is over when a "click" is heard and the Tc-99m button will light up
If all this occurs this initiates the DC is operating normally. If something is wrong an error message should be displayed
Background Correction
Now set the background found in the blue box by select this button
LED should display something close to zero μCi
Let it balance out, close to zero, then select the background button again to disengage this reading
Background button can be selected any time for reset prior to any isotope reading
Complete constancy check with
137
Cs
Place sealed source into ionization chamber
Follow the
lab procedure
as indicated in your lab
Measuring a source of activity
Select the corresponding radioisotope button to be measured and place the activity into the ionization chamber
Regarding
very
low amounts of activity
Prior to placing the activity into the chamber look at the LED display, watch it "swing" back and forth as it registers background
Select the highest reading. It should be in μCi (A)
Now place your source in the chamber and again look for the highest reading (B)
Apply the following to determine net activity B - A = net activity
Selecting a radioisotope that is not one of the buttons - Example
60
Co
Select the "Other" button in the purple box
The press "setup" button 4 times (yellow box)
The LED above the button lights up and flashes
Key in the correct number that correspond to the
Radioisotope Setting
For
60
Co the value is 5.0
Using the up/down arrow move the reading to 5.0
Then hit the Enter key
The "other" button is now set to for
60
Co
The above are different types of dose calibrators
Return to the Table of Content
10/16