- First look at the figure above, which is an example of an electrocardiogram (EKG)
- The following parts of the EKG wave or electrical activity of the myocardium are identified:
P - signal for sinus node impulse to depolarize the atria
QRS complex - electrical depolarization and contraction of the ventricles
P - recovery phase of the ventricles
- Three leads are attached to the patient in order to pick up the R to R interval which is needed for a MUGA acquisition
- How is the heart gated?
- The EKG signal is gated, from each R to R interval
- From the above diagram notice how the R to R interval is subdivided into 5 sections
- Each section, frame, or gate is considered as a segment of time in which counts are collected and stored
- Each time the computer records an R wave acquisition is reset to the first frame and more counts are stored within each individual frame
- Acquisition continues until enough counts are collected in each frame
- Images above the EKG wave represent the data collected in each frame
- In a resting image usually 400 beats or R waves collected is considered sufficient
- Images can then be played back in a dynamic or cine mode that identifies myocardial wall motility (see image below - "Displaying the MUGA")
Note: For demonstration purpose only, 5 gates were used in the above diagram. Usually, the amount of gates set from the R to R interval is as few as 16 or as many 32.
- Wall motion
- Images show wall motion of the heart, however, you are actually seeing labeled RBCs moving in and out of the heart chambers
- There are four types of wall motion to consider:
Normal contractility of all walls
Hypokinesis - one wall or section is moving slower than the rest
Akinesis - one wall or section is not moving at all when compared to the rest
Dyskinesis - one wall or section is moving in opposite direction to the other
- Angles that are imaged
- ANT
- RAO at 10 to 15 degrees
- LAO 30 - 45% (look for angle that best separates the LV)
- Steep LAO or LEFT LAT